Operetta CLS: Unterschied zwischen den Versionen

Aus Imaging Techniken
Keine Bearbeitungszusammenfassung
Keine Bearbeitungszusammenfassung
 
Zeile 26: Zeile 26:
|Objectives: ||  
|Objectives: ||  
:3 air and 3 water-emmersion objectives permanently installed
:3 air and 3 water-emmersion objectives permanently installed
:All objectives in the following table except 20x hNA air and 40x hNA air
:All objectives in the table below
|-
|-
|Laser: ||  
|Laser: ||  
Zeile 34: Zeile 34:
:Harmony 4.9
:Harmony 4.9
|}
|}
{| class="wikitable" style="margin:0"
|+ style="text-align:left;" |<span style="font-variant:small-caps">Objectivs</span>
|-
! style="width:100px; text-align:left;" |Objective !! style="width:80px; text-align:left;" |NA !! style="width:100px; text-align:left;" |Working distance [mm] !! style="width:150px; text-align:left;" |Corrected for plate thickness [mm] !! style="width:120px; text-align:left;" |Field of view [mm] !! style="width:90px; text-align:left;" |Dept of focus [µm]
|-
| 1.25x Air || 0.03 || 4 || 0.17 || 10.33 x 10.33 || 1501.0
|-
| 5x Air || 0.16 || 12.1 || 0.17 || 2.58 x 2.58 || 58.2
|-
| 10x Air || 0.3 || 5.2 || 0.17 || 1.29 x 1.29 || 16.2
|-
| 20x Air || 0.4 || 8.28 || 0 - 1.5 || 0.65 x 0.65 || 8.0
|-
| 20x Water || 1.0 || 1.7 || 0.17 || 0.65 x 0.65 || 1.8
|-
| 40x Air || 0.6 || 3.28 || 0 - 1.5 || 0.32 x 0.32 || 3.3
|-
| 40x Water || 1.1 || 0.62 || 0.17 || 0.32 x 0.32 || 1.2
|-
| 63x Water || 1.15 || 0.6 || 0.17 || 0.21 x 0.21 || 1.0
|}





Aktuelle Version vom 7. Mai 2025, 10:59 Uhr

Type of device

High-Content Imager / Confocal Scanning Microscopy / Live-Cell Imaging
Operetta CLS by PerkinElmer (Revity)


Short description

The device is flexible for different imaging situations that require quantification:

  • Well plate plate types and slides
  • Automated re-imaging, e.g., re-image cells with 10% largest nuclei at higher magnification
  • Widefield and spinning-disk confocal imaging
  • Live-cell imaging (37°C, CO2 variable) over time also with water-objectives (autom. refill)


Instrument parameters

Technical specifications:
sCMOS camera
Motorized stage
Confocal Disk
Transmitted light and digital phase contrast
Ratiometric FRET
3D-imaging and live cell imaging
Fluorescence, fluorescence transmittance, transmission
Objectives:
3 air and 3 water-emmersion objectives permanently installed
All objectives in the table below
Laser:
8 LEDs
Software:
Harmony 4.9


Objectivs
Objective  NA  Working distance [mm] Corrected for plate thickness [mm] Field of view [mm] Dept of focus [µm]
1.25x Air 0.03 4 0.17 10.33 x 10.33 1501.0
5x Air 0.16 12.1 0.17 2.58 x 2.58 58.2
10x Air 0.3 5.2 0.17 1.29 x 1.29 16.2
20x Air 0.4 8.28 0 - 1.5 0.65 x 0.65 8.0
20x Water 1.0 1.7 0.17 0.65 x 0.65 1.8
40x Air 0.6 3.28 0 - 1.5 0.32 x 0.32 3.3
40x Water 1.1 0.62 0.17 0.32 x 0.32 1.2
63x Water 1.15 0.6 0.17 0.21 x 0.21 1.0


Laser: 8 LEDs
Wavelength  Power  Excitation Filter Classification
365 nm (UV) 130 mW 355-385 nm Risk Group 3
405 nm 230 mW 390-420 nm Risk Group 3
450 nm 230 mW 435-460 nm Risk Group 3
475 nm 110 mW 460-490 nm Risk Group 2
505 nm 100 mW 490-515 nm Risk Group 2
550 nm 170 mW 530-560 nm Risk Group 2
630 nm 120 mW 615-645 nm Risk Group 2
660 nm 250 mW 650-675 nm Risk Group 3


Autofocus Laser
Wavelength  Laser Power  Descrption
785+/-10 nm 12 mW

sCMOS camera
Laser class 3B


Transmission Light Source
Wavelength  Laser Power  Descrption
740 nm 80 mW

LED for Brightfiled and Digital Phase Contrast measurements
Integrated in sample lid



Methods & Expertise

Many different imaging assays and quantifications, e.g., migration assay, 3D spheroid growth tracking, organoid imaging and three-dimensional object segmentation and quantification, zebrafish larvae imaging, actin length quantification, scratch assays, single-cell sorting verification (single cell per well) > all in 96-well plates or similar; whole tissue section imaging on up to four slides, prescan-rescan assays for automated re-imaging of specific sub-populations within cell lines and primary cells


Reference publications

Over 10 publications together with University (Medicine) Greifswald, e.g., on human organoids (DOI: 10.1007/s00018-022-04355-6 Fig. 4e)
Over 30 publications in house, e.g., on high-throughput micronucleus analysis (DOI: 10.3390/nano12244463 Fig. 1), actin protrusion quantification (DOI: 10.1038/s41598-018-37169-3 Fig. 5), and whole-tissue imaging and quantification (DOI: 10.1080/2162402X.2024.2425124 Fig. 1-3)


Affiliation, Contact & Terms of use

Affiliation:
XXXX
Location:
XXXX
Contact person:
XXXX
Terms of use:
XXXX
Biosafety level:
None/S2